Power Failure

In the Event of a Power Failure, Turn Off:

  • all computers
  • steppers - the stage may potentially lose track of position and run away
  • furnaces
    • turn off breakers for the individual furnaces
    • open the side doors for ventilation
  • fume hoods - leave the sashes open! Evacuate the wet lab.
  • air handlers - no action, they will be taken care of by F&S

If the power is still out after 15 minutes, students will be dismissed.

Restoration of Power Means Everything Can Be Turned Back On:

but note the following:

  • check the hydrogen detection system to ensure there are no faults
  • e-mail the lab director with a short report of what happened